International Journal of Optics and Applications
p-ISSN: 2168-5053 e-ISSN: 2168-5061
2012; 2(6): 80-86
doi: 10.5923/j.optics.20120206.01
Daniele De Luca1, Ines Delfino2, Maria Lepore1, 3
1Dipartimento di Medicina Sperimentale, Seconda Università di Napoli, Napoli, I-80100, Italy
2Dipartimento Scienze Ecologiche e Biologiche, Università della Tuscia, Viterbo, I-01100, Italy
3CNISM, Unità di Napoli SUN, Napoli, I-80100, Italy
Correspondence to: Ines Delfino, Dipartimento Scienze Ecologiche e Biologiche, Università della Tuscia, Viterbo, I-01100, Italy.
Email: |
Copyright © 2012 Scientific & Academic Publishing. All Rights Reserved.
Laser sources are nowadays largely adopted in medicine and hence they are widespread in medical environment, where patients are present and the users are not always highly specialized in managing laser sources. This has greatly boosted the attention towards safety issues related to exposure to laser beams and to strictly assess the values of well defined laser radiation standard parameters characterizing the level of hazard of laser sources. In this framework, we measured two of the most important parameters, Maximum Permissible Exposure (MPE) and Nominal Ocular Hazard Distance (NOHD), for some of the laser sources mostly employed in medicine. Additionally, we compared our results with data elaborated from standards in order to single out safe and comfortable working conditions. The results here reported have shown that information provided by manufacturers is often not enough to define the hazard level of the laser source and measurements of the main safety parameters are mandatory.
Keywords: Laser, Medicine, Laser Radiation Safety Standards, Maximum Permissible Exposure, Nominal Ocular Hazard Distance
Cite this paper: Daniele De Luca, Ines Delfino, Maria Lepore, "Laser Safety Standards and Measurements of Hazard Parameters for Medical Lasers", International Journal of Optics and Applications, Vol. 2 No. 6, 2012, pp. 80-86. doi: 10.5923/j.optics.20120206.01.
(1) |
(2) |
(3) |
|
Figure 1. Experimental apparatus for irradiance measurements. D: Measuring distance; A: Diaphragm diameter aperture |
|
(4) |
|
[1] | Shaojun Guo, Shaojun Dong, “Graphene nanosheet: synthesis, molecular engineering, thin film, hybrids, and energy and analytical applications”, Chemical Society Reviews, vol.40, no.5, pp.2644-2672, 2011. |
[2] | Carlo Camerlingo, Ines Delfino, Maria Lepore, “Micro-Raman spectroscopy on YBCO films during heat treatment”, Superconductor Science and Technology, vol. 15, no. 11, pp.1606-1609, 2002. |
[3] | Ines Delfino, Anna Rita Bizzarri, Salvatore Cannistraro, “Time-dependent study of single-molecule SERS signal from yeast cytochrome c”, Chemical Physics, vol.326, no.2-3, pp.356-362, 2006. |
[4] | Carlo Camerlingo, Flora Zenone, Ines Delfino, Nadia Diano, Damiano Gustavo Mita, Maria Lepore, “Investigation on clarified fruit juice by using visibile light micro-Raman spectroscopy”, Sensors, vol.7, pp.2049-2061, 2007. |
[5] | Evgeni Eltzov, Serge Cosnier, Robert S Marks, “Biosensors based on combined optical and electrochemical transduction for molecular diagnostics”, Expert Review of Molecular Diagnostics, vol.11, no.5, pp.533-546, 2011. |
[6] | Laser in Dentistry, Proceedings of SPIE vol. 5687, 2005, and references therein. |
[7] | Carlo Camerlingo, Ines Delfino, Giuseppe Perna, Vito Capozzi, Maria Lepore, “Micro-Raman spectroscopy and univariate analysis for monitoring disease follow-up”, Sensors, vol.11, no.9, pp.8309-8322, 2011. |
[8] | Carlo Camerlingo, Maria Lepore, Giovanni M. Gaeta, Roberto Riccio, Carlo Riccio, Alfredo De Rosa, Mario De Rosa, “Er:YAG laser treatments on dentine surface: Micro-Raman spectroscopy and SEM analysis”, Journal of Dentistry, vol.32, no.5, pp.399-405, 2004. |
[9] | Peter Rechmann, Dan S. Goldin, Thomas Hennig, “Er-YAG laser in dentistry: an overview” in Proceedings of SPIE, vol. 3248, pp.2-13, 1998. |
[10] | Susan Walgrave, M. Amanda Jacobs, David Kist, HTL (ASCP), Amy L. Weaver, Elsie Weiler, Irmina Wallander, Brian Zelickson, “Survey of Regional Laser Centers: A Minnesota Perspective”, Dermatological Surgery vol. 37, pp.612–618, 2011. |
[11] | Thomas G, Ash C, Hugtenburg R, Kiernan M, Town G, Clement M., “Investigation and development of a measurement technique for the spatial energy distribution of home-use intense pulsed light (IPL) systems”, Journal Medical Engineering & Technology, vol.35, pp. 191–196, 2011. |
[12] | Yang Jiaqi, Wang Xinwei,Shi Xiaoguang, Zhou Yan, “Eye-Safety Analysis of Infrared Laser Imaging for Security Surveillance”, Proceeding of SPIE, vol. 8200, pp.82000Z/1-9. |
[13] | EN 60825-1 Standard Part 1: Equipment classification, requirements and user’s guide. European Commitment for Electro-technical Standardization, 1998. |
[14] | EN 60601 – 2 – 22 Standard Part 2: Medical electrical equipments: Particular requirements for the safety of diagnostic and therapeutic laser equipment . European Commitment for Electro-technical Standardization, 1998. |
[15] | A. Mallow, L. Chabot, “Laser safety handbook”, Van Nostrand Reinhold, New York, USA, 1978. |
[16] | ICNIRP, “Revision of guidelines on limit of exposure to laser radiation of wavelengths between 400 nm and 1400 nm”, Health Physics vol.79, No.4, pp.431 – 440, 2000. |
[17] | IEC 60825-1 “Safety of laser products –Part 1: Equipment classification, requirements and user's guide” Ed. 1.2, 2001-08, International Electrotechnical Commission, 3 rue de Varembé Geneva, Switzerland, 2001. |
[18] | Town G, Ash C, Dierickx C, Fritz K, Bjerring P, Haedersdal M., “Guidelines on the safety of light-based home-use hair removal devices from the European Society for Laser Dermatology” Journal of the European Academy of Dermatology and Venereology. vol. 26, no.7, pp.799-811, 2012 and references therein. |
[19] | ANSI Z136.1-2000, “American National Standard for Safe Use of Lasers”, American National Standards Institute, New York, USA, 2000. |
[20] | Adam Husein, “Applications of Lasers in Dentistry: A Review”, Archive of Orofacial Sciences, vol.1, pp.1-4, 2006. |
[21] | Qian Peng, Asta Juzeniene, Jiyao Chen, Lars O Svaasand, Trond Warloe, Karl-Erik Giercksky, Johan Moan, “Lasers in Medicine”, Reports on Progress in Physics, vol.71, no.5, 056701, 2008. |
[22] | Mahiul M.K. Muqit, Chintan Sanghvi, Rita McLauchlan, Christine Delgado, Lorna B. Young, Stephen J. Charles, George R. Marcellino, Paulo E. Stanga, “Study of clinical applications and safety for Pascal® laser photocoagulation in retinal vascular disorders“, Acta Ophthalmologica, vol.90, pp.155–161, 2012. |
[23] | Ernst Sutter, Alfred Schirmarcher, “Protective area of laser eye protectors”, Optics & Lasers Technology, vol.33, no.4, pp.255-258, 2001. |