International Journal of Instrumentation Science

p-ISSN: 2324-9994    e-ISSN: 2324-9986

2013;  2(1): 1-5

doi:10.5923/j.instrument.20130201.01

Analysis of Detector-Source Distance and Detector Bias Voltage for Time Resolution of Plastic Scintillation Detectors

E. E. Ermis, C. Celiktas

Ege University, Faculty of Science, Physics Department, 35100, Bornova, Izmir, Turkey

Correspondence to: E. E. Ermis, Ege University, Faculty of Science, Physics Department, 35100, Bornova, Izmir, Turkey.

Email:

Copyright © 2012 Scientific & Academic Publishing. All Rights Reserved.

Abstract

The effects of the source-detector distance, the detector bias voltage variation and collimator usage on the time spectra of 133Ba and 207Bi calibration sources were investigated through a spectrometer which was consisted of a BC400 type general purpose plastic scintillation detector. Constant fraction timing method was used to test these effects on the time resolution values. Optimum source-detector distance and bias voltage values with and without collimator were determined for the optimum time resolution. It was concluded that the source to detector apart, the applied bias voltage to the used detector and the collimator usage were quiet effective parameters in coincidence timing experiments.

Keywords: 133Ba, 207Bi, Constant Fraction Timing Method, Time Resolution

Cite this paper: E. E. Ermis, C. Celiktas, Analysis of Detector-Source Distance and Detector Bias Voltage for Time Resolution of Plastic Scintillation Detectors, International Journal of Instrumentation Science, Vol. 2 No. 1, 2013, pp. 1-5. doi: 10.5923/j.instrument.20130201.01.

1. Introduction

Timing measurement focuses on the determination of the time when a particular event has happened (i.e., a charged particle has crossed the detector), with a certain accuracy. The information on the energy associated with the event has a lesser priority or may even be disregarded. This fact fundamentally changes the way timing radiation systems are designed and optimized[1].
Several different nuclear radiation detectors are used in timing experiments: scintillation counters with organic and NaI(Tl) scintillators, Cherenkov detectors, multiwireproportional chambers, semiconductor detectors. Scintillation counters with organic scintillators, however, are still superior to produce the best time resolution, as well as to be used in a variety of different applications.
The time resolution of scintillation counters is generally determined by the method used in the so-called delayed-coincidence technique which is widely applicable for the measurements of short nuclear life-times. The time resolution of a timing system can be determined by measuring the distribution of the time delays between prompt events.
The figure of merit of a system is usually characterized by the full width at half maximum (FWHM)[2].
The premium plastic scintillators such as BC400 are the most economical as well as their highest light output[3]. The most common and widely used plastics are polyvinyltoluene, polyphenylbenzene and polystyrene. Plastics offer an extremely fast signal with a decay constant of about 2-3 ns. Because of this fast decay, the finite rise time cannot be ignored in the description of the light pulse[4].
Using NaI(Tl) scintillation detector, time spectrum of 22Na was obtained by Nutt et al.[5]. Time spectrum of mixed radioactive alpha source (239Pu, 241Am and 244Cm) was given by means of multichannel plates (MCPs) and Si detectors by Sharma et al.[6]. Kawada et al.[7] presented time spectra of 60Co at different temperatures. Optimization of time resolution of a general purpose plastic scintillation detector through leading edge timing method was investigated by Ermis and Celiktas[8]. Ermis and Celiktas[9] carried out a study about time resolution investigations for general purpose plastic scintillation detectors in different thicknesses. Parnham and Eissler[10] depicted the time spectrum of 22Na radioisotope via a CdZnTe detector.
Since the time resolution is a critical parameter for timing measurement setups, a spectrometer was constructed by using a BC400 type plastic scintillation detector. Time spectra of 133Ba and 207Bi radioisotopes were obtained for the determination of best time resolution by means of constant fraction timing technique. Optimization of time resolution of this type detector was carried out by depending on the source-detector distance and the detector bias voltage variation together with collimator effect.

2. Experiments

133Ba and 207Bi calibration sources with the activities of 3.5 x105 Bq were used in this work. Their half-lives are 10.51 and 31.55 y, respectively[11]. The thicknesses and the diameters of the sources were 0.3 and 2.5 cm, respectively. The sources were placed in various distances far away from the detector active surface, perpendicularly to the detector normal direction.
A Bicron BC400 type (equivalent to NE102A) plastic scintillation detector with 3 mm thickness and 3 inch diameter was used because this type plastic scintillator is preferred for general purpose radiation measurement applications[3].
Constant fraction timing method, which is one of the timing methods, was used in this work. Probably the most efficient and versatile method available today is the constant fraction triggering technique[4, 12, 13]. In this method, a logic signal is generated at a constant fraction of the peak height to produce an essentially walk-free timing signal. The basis for this idea arose from empirical tests which showed the existence of an optimum triggering level for the best timing resolution. Depending on the type of signal, this level occurs at a certain fraction of the pulse height independent of the amplitude[4]. The principle of constant-fraction timing can be found in[4, 12, 13].
Block diagram of the used spectrometer is given in Fig. 1.
Figure 1. Block diagram of the used spectrometer
Figure 2. Signal shapes of (a) TFA and (b) TAC outputs
In the setup, the detector output was split into two branches, and these outputs were forwarded to the timing filter amplifiers (TFA, Ortec 454). The outputs of the TFAs were connected to the constant fraction discriminators (CFD, Ortec 463). One of the output signals of the CFDs was directly sent to the time to amplitude converter (TAC, Ortec 566) and the other output was connected to the TAC through a delay unit (D, Ortec 425A). The output signal shapes from the TFA and the TAC can be seen in Fig.2.
The experimenter must choose a time constant that is at least 5-10 times greater than the scintillator decay time (2.4 ns) but that is not excessively long to prevent needless pulse pileup with the tail from a preceding pulse at high rates[13]. Therefore, the time constants of the TFAs were chosen as 50 ns.
The time to amplitude converter (TAC) is a device that produces an output pulse with an amplitude proportional to the time interval between input ‘start’ and ‘stop’ pulses. The differential amplitude distribution of the output pulses as recorded by a multichannel analyzer is thus a measure of the distribution of time intervals between start and stop pulses and is often called a time spectrum[13]. Optimum delay time for the stop signal of TAC was 16 ns. TAC range was chosen as 200 ns because optimum time signals were obtained at this range. Time spectra of the radioisotopes were recorded in a computer via a multichannel analyzer (MCA- Ortec trump 8K).
Timing in nuclear and particle physics refers to the measurement of very small time intervals. Accurate measurements of very small time intervals require special techniques. The most important factor in any timing system is its resolution, i.e., the smallest time interval that can be measured with accuracy[4].
Before acquiring the time resolution results, we first calibrated the channels of the MCA, and we plotted a time calibration graph for this purpose as shown in Fig. 3. The inverse of the slope of this fitted graph gives approximately time value (195.3125 ± 0.0051 ps) of each channel of the MCA. This value may calculate by dividing the range of the TAC to the conversion gain of the MCA. As a result, the calibration values obtained from both ways were almost equal. This showed that good time calibration was acquired.
The time spectra of the radioisotopes were collected by depending on the source-detector distance and the applied bias voltage variation. In addition, the effect of the collimator usage on the results was observed. An Al collimator was used with a thickness of 1 cm. The collimator was mounted on the active surface of the sources. The hole diameter and the outer diameter of the collimator were 0.5 cm and 2.5 cm, respectively (Fig. 1).
Laboratory conditions such as ambient light intensity and increase in the room temperature result in detector noise (dark noise), leading to additional counts[4]. Therefore, all measurements were acquired at a constant room temperature (20 ℃) and in a dark medium with a data acquisition time of 300 s.
Figure 3. Time calibration graph (fitted time calibration=195.3125 ± 0.0051 ps/ch)

3. Results

To find out optimum time resolution according to the variation of detector-source apart, we collected time spectra of the sources at various detector-source distances without and with collimator. Best time resolution values were obtained at 6 cm from the measurements as given in Table 1.
Since optimum time resolution results were obtained in the detector-source apart of 6 cm for the used radioisotopes without and with collimator, they were placed in turn at 6 cm from the detector window for the optimization of detector bias voltage at this distance. We changed the applied voltage to the detector from 900 V to 1,300 V. It was observed that the time resolution values were saturated at 1,200 V-1,300 V as summarized in Table 2.
Table 1. Time resolution results with respect to the source-detector distance
     
Table 2. Time resolution results with respect to the detector bias voltage
     
It can be deduced from these results that the optimum bias voltage for the detector was 1,200 V as recommended by the manufacturer. The results did not available at first lines in Table 2 since we could not obtain any spectrum of 133Ba without and with collimator at 900 V.
If the dead time obtained from the MCA exceeds 30-40% level, pulse pile-up and spectrum distortion will occur. Piled-up pulses broaden the time peak and deteriorate the resolution[4, 13]. Since the dead time from the MCA was around 1% in the present study, it can be concluded that our results did not contain these malicious effects.

4. Conclusions

The effects of the source-detector distance and the detector bias voltage variation on the time spectra of 133Ba and 207Bi calibration sources were investigated through a spectrometer consisted of a BC400 type plastic scintillation detector by means of constant fraction timing method. Optimum source-detector distance and bias voltage values were determined for the best time resolution.
Experimental results showed that the optimum time resolution for both radioisotopes without and with collimator were obtained when the source-detector distance was set to 6 cm as shown in Table 1. It can be deduced from this result that the detector to source distance is quite effective on time resolution. This result points out that the effect of the detector-source distance on time resolution should be taken into consideration especially in coincidence measurements.
Effect of the variation of the detector bias voltage to the time resolution was also investigated. In order to analyze this effect, both radioisotopes without and with collimator were placed in turn at 6 cm far away from the detector surface. The time resolution was high at low detector bias voltage, leading to poor resolution. It is noticeable that as the bias voltage increases the resolution value decreases until 1,200 V level, leading to better resolution. It was deduced from the results that time resolution has saturated at the applied voltage of 1,200 V-1,300 V as depicted in Table 2. Optimum detector bias voltage for the best time resolution was found so as to be 1,200 V. This result proves us that the best time resolution is obtained at the recommended detector bias voltage by the manufacturer. In addition, positive effect of the collimator usage on time resolution can be noticed from the tables above.
Consequently, it was pointed out from the present study that the source to detector apart, the applied bias voltage to the used detector and the collimator usage were quiet important parameters in coincidence timing experiments. Collimator usage, optimum source-detector apart and bias voltage should be determined before data acquisition.

ACKNOWLEDGEMENTS

This work was supported by TUBITAK, the Scientific and Technological Research Council of TURKEY under Project No. 197T087 and by EBILTEM, Center of Science and Technology, Ege University under Project No. 99 BIL 001

References

[1]  S. Martoiu and A. Rivetti, ‘Electronics for Radiation Detection’, CRC Press, USA, 2011.
[2]  M. Moszynski M and B. Bengston, ‘Status of Timing with Plastic Scintillation Detectors’, Nucl. Instrum. and Meth., vol. 158, no. 1979, pp. 1-31, 1979.
[3]  Saint-Gobain Scintillation Products, ‘Premium Plastic Scintillators’,http://www.detectors.saint-gobain.com/uploadedFiles/SGdetectors/Documents/Product_Data_Sheets/BC400-404-408-412-416-Data-Sheet.pdf , Accessed 29 November 2012.
[4]  R.W. Leo, ‘Techniques for Nuclear and Particle Physics Experiments’, Springer–Verlag Berlin Heidelberg, Germany, 1987.
[5]  R. Nutt, D.A. Gedcke, C.W. Williams, ‘A Comparison of Constant Fraction and Leading Edge Timing with NaI(Tl) Detectors’, IEEE Trans. Nucl. Sci. vol. 17, no. 1, pp. 299-306, 1970.
[6]  H. Sharma, K. Kosev, A. Wagner, K.D. Schilling, A. Wagner II, W. Schulze, K. Heidel K, ‘A Time Coincidence Test Setup for an MCP Detector’http://www.hzdr.de/FWK/jb02/teil1.pdf, Accessed 29 November 2012.
[7]  Y. Kawada, J. Ito, Q.W. Wang, ‘Temperature Dependence of Spurious Pulses in Use of Plastic Scintillation Detector’, Appl. Radiat. and Isot., vol. 60, no. 2004, pp. 403-407, 2004.
[8]  E.E. Ermis and C. Celiktas, ‘Effects of Detector-Source Distance and Detector Bias Voltage Variations on Time Resolution of General Purpose Plastic Scintillation Detectors’, Appl. Radiat. and Isot., vol. 70, no. 12, pp. 2682-2685, 2012.
[9]  E.E. Ermis and C. Celiktas, ‘Time Resolution Investigations for Genaral Purpose Plastic Scintillation Detectors in Different Thickness’, J. Radioanal. and Nucl. Chem., in press, 2012.
[10]  K.B. Parnham and E.E. Eissler, ‘A Study of the Timing Properties of Cd0.9Zn0.1Te’, Nucl. Sci Symposium and Medical Imaging Conference Record, vol. 1, pp. 136-138, 1995.
[11]  Table of Radioactive Isotopes, http://ie.lbl.gov/toi/, Accessed 29 November 2012.
[12]  N. Tsoulfanidis, ‘Measurements and Detection of Radiation’, Taylor&Francais, USA, 1995.
[13]  G.F. Knoll, ‘Radiation Detection and Measurements’, John& Sons. Inc, New York, 2000.